Techne TC-3000X Manuel d'utilisateur Page 34

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 55
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 33
EXAMPLE PROGRAMS
A couple of example thermal cycling programs are shown below.
Typical 3-Step PCR 3-Step PCR with multiple 'Stages'
32
PROGRAMMING
To add a stage and then edit it
Heated Lid 105°C
Preheat Lid on
Pause off
-------------------------
In denat 94°C 5min
Hot Start off
-------- Stage 1 --------
Number cycles 30
Step 94.0°C 0m30 max
Step 55.0°C 0m30 max
Step 72.0°C 0m30 max
-------------------------
Final Extn 72.0°C 5m
Final hold 4°C 10m
Heated Lid 105°C
Preheat Lid on
Pause off
-------------------------
In denat 94°C 5min
Hot Start off
-------- Stage 1 --------
Number cycles 10
Step 94.0°C 0m30 max
Step 60.0°C 0m30 max
Step 72.0°C 0m30 max
-------------------------
Number cycles 20
Step 94.0°C 0m30 max
Step 55.0°C 0m30 max
Step 72.0°C 0m30 max
-------------------------
Final Extn 72.0°C 5m
Final hold 4°C 10m
Create a new program by selecting ** New
Program ** from the menu and press
"Enter"
Give the program a name and edit any of the
default fields as described above.
Use the "Down Arrow" key to scroll to
the dotted line below 'Hot Start'; it will start
to flash.
Press "Enter" to insert a stage.
* * New Program * * [ ]
2 STEP TEMPLATE 501
3 STEP TEMPLATE 502
[ , to select ]
UNNAMED #
Heated Lid 105°C
Preheat Lid on
[ , to edit ]
In denat 94.0°C 5m00
Hot Start off
-------------------------
[ to insert stage ]
[ to save ]
Vue de la page 33