Techne TC-3000X Manuel d'utilisateur Page 37

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 55
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 36
[ ‘.’ to insert step ]
[ , to edit ]
TO PROGRAM INCREMENT OR DECREMENT TEMPERATURE OR TIME
Under normal circumstances, the Hold Temperature of all steps is constant. However, it is possible to
automatically increment or decrement the temperature of a specified step of a programme.
Temperature profile of incremented Hold Temperature at step 3 of each cycle
35
To insert a pause after an existing stage,
follow the steps for inserting a stage as
above.
Highlight the 'Number cycles' line then press
"Pause"
The prompt line will flash between 'CHANGE
TO PAUSE?' and 'YES or NO'.
Press "Enter" to insert a pause.
The stage number will then be highlighted
with 'Pause' as the only line.
To insert a ‘pause’ into a program
Step 94.0° C 00m30 max
-------- Stage 2 --------
Number cycles
Step 94.0° C 00m30 max
[ CHANGE TO PAUSE ]
[ Yes or No ]
Step 94.0° C 00m30 max
-------- Stage 2 --------
Number cycles 0
Step 94.0° C 00m30 max
-------- Stage 2 --------
Pause
-------------------------
[ to insert step ]
[ to delete step ]
Vue de la page 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 54 55

Commentaires sur ces manuels

Pas de commentaire