Techne TC-3000G Manuel d'utilisateur Page 22

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 56
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 21
HEATED LID
The heated lid has a rotating knob to adjust the lid
pressure on the samples, allowing for a variety of
consumables to be used. To adjust the pressure:
• Adjust the orange rotating knob anticlockwise
to give the least pressure.
• Place the samples in the block and place the lid
in the down position.
• Gently rotate the knob clockwise until you can
feel pressure beginning to be applied.
• Rotate the knob a further quarter of a turn and
this will give the correct pressure.
20
USER OPERATION AND CONSUMABLES
Do not over-tighten the knob as this will cause deformation of the silicone rubber mat (where used) and
reduce the effectiveness of the lid. Once the pressure has been set it does not need to be adjusted unless
a different block or consumable is used.
The silicone rubber mat supplied with the unit (Part number 6106196) is for use with 48 well microplates
when sealing using a plastic adhesive film. The mat should be placed on top of the sealing film to ensure
good thermal contact with the heated lid.
TUBES OR REACTION VESSELS
Bibby Scientific Ltd does not recommend any specific tube or reaction vessel other than those described
in this manual. We recommend using reaction volumes between 20 and 200µl. The tubes must withstand
a pressure of 1 atmosphere at 100°C. Any vessel must be able to withstand the temperatures you are
using without any danger of them deforming to the point where they fracture.
To test your tubes, put 25µl of water into each of 5 tubes and subject them to a typical thermal cycling
protocol. At the end of the program, measure the volume remaining using a micropipette. A loss of more
than 1 or 2µl indicates a vapour leak.
The amount of volume loss you observe and the change in reactant concentrations you can tolerate
determine the minimum volume that can be used. Typical volume losses of 1µl in 30 cycles allow the use
of samples of 20µl or less.
During the final cool-down, a ring of condensation may form above the liquid level but below the top of
the sample block. This is normal and does not effect the reaction.
Vue de la page 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 55 56

Commentaires sur ces manuels

Pas de commentaire